Voltage Calibration For Voltage Range Low - Ametek Elgar ReFlex Power Operation Manual

Table of Contents

Advertisement

AC Power Supplies
V
4.6.14
OLTAGE
This procedure calibrates the output voltage of the power supply. This is five-
point calibration procedure.
1. Connect the meter between the output and the output return lead.
2. Set the voltmeter to read the AC voltage.
3. Ensure that no load is connected to the output.
4. Issue clear and reset commands:
5. Set the voltage range of the power supply to low
6. Set the current to maximum limit
7. Turn the unit on
8. Set the OVP supervisory to limit 20261
9. Set the OCP supervisory to maximum limit.
10. Program the output voltage to 5% of the full scale voltage
11. Wait for AC voltage to stabilize. Measure the output voltage using the AC
voltmeter. Enter the measured voltage using following command
12. Program the output voltage to 27.5% of the full scale voltage
4-60
C
ALIBRATION FOR
*CLS
*RST<n>
SOUR<n>:VOLT:RANGE 0
CAL<n>:OUTP:CURR:COUNTS 4000
OUTP<n>:STATE 1
CAL<n>:OUTP:VOLT:PROT:COUNTS 20261
CAL<n>:OUTP:CURR:PROT:COUNTS 4000
CAL<n>:OUTP:VOLT:COUNTS 965
CAL<n>:OUTP:VOLT:FIVEPOINT1 <value-1>
CAL<n>:OUTP:VOLT:COUNTS 5307
V
R
OLTAGE
ANGE
ReFlex Power™
L
OW
M380056-01 Rev L