Hall Calibration For Voltage Range Low; All Alibration For Oltage Ange Ow (5 Kh Nit ) - Ametek Elgar ReFlex Power Operation Manual

Table of Contents

Advertisement

ReFlex Power™
H
4.6.20
ALL
U
NIT
The following procedure calibrates the hall gain sensor of the power supply for
the 5KHz AC unit only. This is seven-point calibration procedure.
1. Close the switch between the power supply and the resistive load.
2. Connect the meter between the output and adjustable resistive load.
3. Set the voltmeter to read the AC current.
4. Ensure that a resistive load is connected to the output.
5. Issue clear and reset commands:
6. Set the voltage range of the power supply to low:
7. Turn the unit on:
8. Set the hall calibration data to the default value:
9. Set the output voltage to 120 Volts:
10. Set the output current to 6.3 Amps:
11. Turn on the resistive load and adjust the load until the output voltage is 110 Volts,
this will ensure that the AC supply is in Constant Current Mode.
12. Set the output frequency to 60Hz
13. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
14. Set the output frequency to 500Hz
M380056-01 Rev L
C
ALIBRATION FOR
)
*CLS
*RST<n>
SOUR<n>:VOLT:RANGE 0
OUTP<n>:STATE 1
CAL<n>:OUTP:FREQ:CURR:DEFAULT
SOUR<n>:VOLT 120
SOUR<n>:CURR 6.3
SOUR<n>:FREQ 60
CAL<n>:OUTP:FREQ:CURR1 <value-1>
SOUR<n>:FREQ 500
V
R
OLTAGE
ANGE
AC Power Supplies
L
(5 KH
OW
Z
4-73