Lpal Resistance Calibration - Part Iv Ofv - Ametek Elgar ReFlex Power Operation Manual

Table of Contents

Advertisement

Active Loads
LPAL R
7.6.30
This procedure calibrates the resistance for voltage gain 1 and current
range high. Make sure that the shunt and the DC voltmeter are connected
between the positive and negative lead of the active load. Connect the
sense positive-to-positive lead and sense negative-to-negative lead. Also
connect the shunt and the DC current meter between the active load and
DC power supply
This procedure requires maximum voltage 100.0 V and maximum current
0.5 Amps.
1. Set the DC input voltage to 100 Volts
2. Turn the active load
3. Set the conductance count to 0xC7C.
4. After DC voltage stabilizes. Enter the voltage using SCPI
5. After DC current stabilizes. Enter the current using SCPI commands
6. Set the conductance count to 0x4EC.
7-78
C
ESISTANCE
ALIBRATION
INP<n>:STATE 1
CAL<n>:INP:COND:COUNTS 0xC7C
commands
CAL<n>:INP:VOLT 1 <value-1>
CAL<n>:INP:CURR 1 <value-1>
CAL<n>:INP:COND:COUNTS 0x4EC
– P
IV
V
ART
OF
ReFlex Power™
M380056-01 Rev L