Active Loads
7. This will use the current value entered to adjust the Gain Pot. The
8. Send the system error query command to make sure calibration was
9. If there is no error returned at the system error query command then
10. Optional Commands:
11. Reset the unit
7-56
adjustment will be calculated, ending the procedure and the setting
the load to STATE 0
CAL<n>:CALPOTS:GAIN:CALC
successful.
SYST:ERR?
save the calibration result with the following command
CAL<n>:UNLOCK "6867"
CAL<n>:STORE
CAL<n>:LOCK
To query the offset pot set value: CAL<n>:CALPOTS:OFFSET?
To query the gain pot set value: CAL<n>:CALPOTS:GAIN?
*RST<n>
ReFlex Power™
M380056-01 Rev L