DC Power Supplies
10. Measure the output current with the external shunt. Enter this as
11. Program the output current to the third value near 50% of full-scale,
12. Measure the output current with the external shunt. Enter this as
13. Program the output current to the fourth value near 72.5% of full-scale,
14. Measure the output current with the external shunt. Enter this as
15. Program the output current to the fifth value near 95% of full-scale,
16. Measure the output current with the external shunt. Enter this as
17. Save the calibration results with the following commands:
6-70
(count = 32768*0.25256 = 8276):
CAL[n]:OUTP:CURR:COUNTS 8276
value-2 with the following command:
CAL[n]:OUTP:CURR:FIVEPOINT 2 <value-2>
(count = 32768*0.46219 = 15145):
CAL[n]:OUTP:CURR:COUNTS 15145
value-3 with the following command:
CAL[n]:OUTP:CURR:FIVEPOINT 3 <value-3>
(count = 32768*0.67258 = 22039):
CAL[n]:OUTP:CURR:COUNTS 22039
value-4 with the following command:
CAL[n]:OUTP:CURR:FIVEPOINT 4 <value-4>
(count = 32768*0.88132 = 28879):
CAL[n]:OUTP:CURR:COUNTS 28879
value-5 with the following command:
CAL[n]:OUTP:CURR:FIVEPOINT 5 <value-5>
CAL[n]:UNLOCK "6867"
CAL[n]:STORE
CAL[n]:LOCK
(The entered current data could be reviewed with the following query:
CAL[n]:OUTP:CURR:FIVEPOINT?)
ReFlex Power™
M380056-01 Rev L