Hpal Voltage Calibration - Part Vofv - Ametek Elgar ReFlex Power Operation Manual

Table of Contents

Advertisement

Active Loads
HPAL V
7.6.12
This procedure calibrates the voltage for voltage gain 1 and current range low.
Make sure that the shunt and the DC voltmeter are connected between the
positive and negative lead of the active load. Also connect the sense positive-to-
positive lead and sense negative-to-negative lead.
This procedure requires maximum voltage 500V and maximum current 0.18
Amps.
1. Turn the active load on:
2. Set the conductance count to 0x15D:
3. Set the input DC voltage to 10 Volts. After the input DC voltage stabilizes
4. Set the input DC voltage to 500 Volts. After the input DC voltage
5. Send the calibration command to calculate the gain and offset:
6. Send the system error query command to make sure calibration was
7. If there is no error returned at the system error query command then save
8. Reset the unit:
7-42
C
OLTAGE
ALIBRATION
INP<n>:STATE 1
CAL<n>:INP:COND:COUNTS 0x15D
enter the first exact voltage using following command:
CAL<n>:INP:VOLT 1 <value-1>
stabilizes enter the second exact voltage using following command:
CAL<n>:INP:VOLT 2 <value-2>
CAL<n>:MEAS:OFFSET:CALC
successful:
SYST:ERR?
the calibration result:
CAL<n>:UNLOCK "6867"
CAL<n>:STORE
CAL<n>:LOCK
*RST<n>
– P
V
V
ART
OF
ReFlex Power™
M380056-01 Rev L