System Settings (Menu); Leveling The Pipette Calibration System (Q-Level) - Sartorius cubis MPS Quick Reference Manual

Electronic pipette calibration system
Table of Contents

Advertisement

Please select the menu item:
Menu
Level balance
Calibration/adjustment data
Configure timer controlled actions
Display device information
Device parameters
Back
Language
4

System Settings (Menu)

This menu contains all settings relevant to the device. Basic settings can be made
here that immediately affect the device. Changed settings do not affect previously
defined tasks and user profiles.
All system settings are user-independent; in other words, they are applied to all users.
As long as a task is active and/or a user has been selected, several basic settings
cannot be changed. In order to enable all settings to be changed, "Weighing" has
to be selected in the operating mode and no user profile may be selected.
t
In order to open the system settings, touch
y
The selection of the system settings is displayed.
t
To view all menu items, use the right scroll bar.
t
Touch the menu item whose settings you want to change.
System settings are divided into the following menus:
− Leveling the pipette calibration system (see next section, page 15)
− Configuring calibration/adjustment (see page 17)
− Setting up timer-controlled actions (see page 18)
− Displaying device information (see page 21)
− Configuring the device (see page 22)
− Importing/exporting data (see page 29)
− Activating the service mode (see page 32)
You have already learned how to use the menus. In the following, only configuration
options are described and explained.
4.1

Leveling the Pipette Calibration System (Q-Level)

The pipette calibration system must be leveled in order to enable it to provide
accurate weighing results. During the leveling process, the tilt of a weighing plate
is eliminated by adjusting the leveling feet of the pipette calibration system. Leveling
is always required if the pipette calibration system is set up at a different location or
is moved at its original location.
The pipette calibration system has leveling feet that have to be manually adjusted.
An integrated sensor detects the alignment of the pipette calibration system and
triggers a warning message when leveling is required. You can configure the leveling
notice and leveling (see page 23).
The pipette calibration system needs to be re-leveled and then adjusted each time its
setup location is changed.
System Settings (Menu)
in the operating mode.
Menu
®
Cubis
MPS Quick Reference Guide
15