Level the balance with the inclination sensor
It is essential for exact weighing results that the balance is absolutely level. The
front leveling feet can be used to level out small tilts of the floor. An integrated
sensor detects the alignment of the balance and triggers a warning message when
leveling is required.
When »
« appears on the display, you must level the balance.
Step
1.
flashes in the display:
Start leveling.
2. Fully screw in both back leveling feet
(depends on the model) and confirm.
3. Directional arrow on the right of TURN:
Turn right leveling foot.
Directional arrow on the left of TURN:
Turn left leveling foot.
Directional arrow
: Turn the respective leveling foot
to the right until the directional arrow disappears.
Directional arrow
: Turn the respective leveling foot
to the left until the directional arrow disappears.
4. Screw out both back leveling feet
(depends on the model) until they touch
the setup surface.
Using the Balance
Press key
Display
flashes
(
)
*
SCREW IN
(
)
TURN
*
TURN
TURN
SCREW OUT
(
)
0.0g
*
Betriebsanleitung Cubis MSE
29