Chapter 12: Emulator Error Messages
Trigger position changed to accomodate trig1, trig2 delay spec (Status 1203)
Trigger position changed to accomodate trig1, trig2 delay spec (Status 1203)
Cause: The terminal interface tgout (trigger output) command provides a delay
feature that allows for driving of the trig1 and/or trig2 signals a specified number of
states after trigger or before trace complete. The setup of this delay feature
interacts with the trigger position specification. The trigger position specification
may be automatically modified by the deep analyzer in order to make the delay
feature work in the expected manner.
Action: You can use the terminal interface command tp (trigger position) to
examine the new trigger position value.
Trigger term cannot be term 1 (Error 1251)
Cause: This error occurs when you attempt to specify the first sequence term as the
trigger term. The trigger term may be any term except the first.
Action: Respecify the trigger term as any other sequence term.
Too many sequence terms (Error 1226)
Cause: This error occurs when you attempt to insert more than four sequence terms.
Action: Do not attempt to insert more than four sequence terms.
Trace error during CMB execute (Error 692)
Cause: System failure.
Action: Run performance verification (pv command).
Trace format command failed; using old format (Error 1133)
Cause: This error occurs when the trace format command fails for some reason.
Action: This error message always occurs with another error message. Refer to the
description for the other error message displayed.
434