Section 2: TSP Programming
Figure 2-3
GPIB cable
NOTE To minimize interference caused by electromagnetic radiation, use
The GPIB cable connectors are stackable. For additional non-Series 2600 GPIB instruments in the
test system, daisy-chain a GPIB cable from one instrument to another.
GPIB address
At the factory, the GPIB is set for address value 26. The address value can be set to any address
value between 0 and 30. However, the address cannot conflict with the address assigned to other
instruments in the system.
The GPIB address can be changed from the communications menu. To access the menu, press
the MENU key, select COMMUNICATIONS, and then select GPIB.
The GPIB address is saved in non-volatile memory. The address value will not change when
power is cycled or a reset command (

RS-232

RS-232 connections
To connect the Series 2600 to the RS-232 interface, use a straight-through RS-232 cable (see
Figure
2-4).
connecting the cable, make sure to tighten the captive screws.
Figure 2-4
RS-232 cable (straight-through)
RS-232 settings
At the factory, the RS-232 is configured as follows:
2-10
only shielded GPIB cables. Available shielded cables from Keithley
Instruments are the Model 7006 and Model 7007.
Figure 2-3
shows the location of the RS-232 connector on the SourceMeter. When
Return to
Section Topics
Series 2600 System SourceMeter® Instruments Reference Manual
or *RST) is sent.
reset
2600S-901-01 Rev. C / January 2008