Master And Slaves; System Configurations; Overview - Keithley 2601 Reference Manual

System sourcemeter 2600 series (smu)
Hide thumbs Also See for 2601:
Table of Contents

Advertisement

Section 9: System Expansion (TSP-Link)

Overview

The Keithley Instruments Series 2600 System SourceMeter
interface that allows the instruments to communicate with each other. The test system can be
expanded to include up to 16 TSP-Link-enabled instruments.

Master and Slaves

In a TSP-Link system, one of the nodes (instruments) is the Master and the other nodes are the
Slaves.
The Master can control the other nodes (Slaves) in the system. When any node transitions from
local operation to remote, it becomes the Master of the system; all other nodes also transition to
remote operation, and become its Slaves. When any node transitions from remote operation to
local, all other nodes also transition to local operation, and the Master/Slave relationship between
nodes is dissolved. For more information about remote and local operations, see
Remote versus local
A Slave is a node that is controlled by the Master. The GPIB and RS-232 command interfaces of
the Slaves are disabled.

System configurations

A TSP-Link system can be used without a PC (stand-alone system) or as a PC-based system.
Stand-alone system – In a stand-alone system, scripts that control the system are executed from
the front panel of one of the instruments. No PC connection is required. In
be run from the front panel of any one of the instruments.
PC-based system – In a PC-based system, the GPIB or RS-232 interface to any single node
becomes the interface to the entire system. In
GPIB or RS-232 interface of Node 1.
Connections
Connections for an expanded system are shown in
connected to the PC using the GPIB or RS-232 interface. Details on these PC communication
connections are covered in
As shown in
cables.
9-2
state" in Section 2.
Section
2.
Figure
9-1, all the units in the system are daisy-chained together using LAN crossover
Return to
Section Topics
Series 2600 System SourceMeter® Instruments Reference Manual
®
TSP-Link™ is an expansion
Figure
9-1, the system can be controlled via the
Figure
9-1. As shown, one unit is optionally
"Differences:
Figure
9-1, a script can
2600S-901-01 Rev. C / January 2008