Series 2600 System SourceMeter® Instruments Reference Manual
Figure 5-5
Custom sweep example

Sweep measurement storage

When sweeps are run, measurements are automatically stored in non-volatile memory Buffer 1 for
later recall. Sweep data can be recalled as follows:
Front panel: Press the RECALL key, select the channel and Buffer 1, then choose reading
numbers to display with the Rotary Knob or cursor keys.
Remote: Use the printbuffer command to request buffer readings from
smua.nvbuffer1 (channel A) or smub.nvbuffer1 (channel B).
See
Section 7

Sweep functions

Functions to perform staircase, pulse, and custom sweeps are discussed below. See
this manual for details on using factory scripts.
NOTE Visit www.keithley.com for additional available user scripts for various
2600S-901-01 Rev. C / January 2008
for details on recalling data from the buffer.
tests.