Power-Up Sequence; Beeper - Keithley 2601 Reference Manual

System sourcemeter 2600 series (smu)
Hide thumbs Also See for 2601:
Table of Contents

Advertisement

Series 2600 System SourceMeter® Instruments Reference Manual

Power-up sequence

On power-up, the Series 2600 performs self-tests on its ROM and NVRAM and momentarily lights
all segments and annunciators. If a failure is detected, the instrument momentarily displays an
error message and the ERR annunciator turns on (error messages are listed in
NOTE If a problem develops while the instrument is under warranty, return it
Assuming no errors occur, the Series 2600 will power-up as follows:
After a few seconds with the OUTPUT indicators and display pixels on, the instrument
model number, firmware revision levels, and line frequency setting are briefly displayed.
The node and the GPIB address are displayed briefly as follows:
1.
KEITHLEY MODEL 26xx
2.
NODE = 1
The node and serial port parameters are displayed briefly:
3.
KEITHLEY MODEL 26xx
4.
NODE = 1
If the line frequency setting is AUTO, a screen will be displayed indicating the power line
frequency is being detected.
System identification
Serial number, firmware revision, and calibration dates can be displayed by selecting the SERIAL#
item of the main menu.
Complete the following steps to view the system information.
1.
Press MENU > SYSTEM-INFO.
2.
Choose one of the following:
For remote programming, use the *IDN? query to read system information.

Beeper

With the beeper enabled, a beep will be issued to acknowledge the following actions:
A short beep, emulating a keyclick, is issued when a front panel key is pressed.
A short beep is also issued when the rotary knob is turned or pressed.
A longer beep is issued when the source output is turned on.
Complete the following steps to turn the beeper on or off.
1.
select MENU > BEEPER.
2.
Choose one of the following:
1. Unit displays actual model number: 2601, 2602, 2611, 2612, 2635, or 2636.
2. Unit displays actual model number: 2601, 2602, 2611, 2612, 2635, or 2636.
2600S-901-01 Rev. C / January 2008
to Keithley Instruments, Inc., for repair.
1
GPIB = 26
2
SERIAL = 9600,8,N,1,NONE
FIRMWARE
SERIAL#
CAL
Return to
Section Topics
Section 1: Getting Started
Appendix
B).
1-15

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

26362602261126122635