Compliance Limit - Keithley 2601 Reference Manual

System sourcemeter 2600 series (smu)
Hide thumbs Also See for 2601:
Table of Contents

Advertisement

Series 2600 System SourceMeter® Instruments Reference Manual
Table 4-1
Source-measure capabilities
Model 2601/2602
Range
Source
Measure
100mV
±101mV
±102mV
1V
±1.01V
±1.02V
6V
±6.06V
±6.12V
40V
±40.4V
±40.8V
100nA
±101nA
±102nA
1µA
±1.01µA
±1.02µA
10µA
±10.1µA
±10.2µA
100µA
±101µA
±102µA
1mA
±1.01mA
±1.02mA
10mA
±10.1mA
±10.2mA
100mA
±101mA
±102mA
1A
±1.01A
±1.02A
3A
±3.03A
±3.06A
Max Power = 40.4W per channel

Compliance limit

When sourcing voltage, the SourceMeter can be set to limit current. Conversely, when sourcing
current, the SourceMeter can be set to limit voltage. The SourceMeter output will not exceed the
compliance limit. The maximum compliance limit is the same as the maximum values listed in
Table
4-2. Note that the compliance value will take the same sign as the source value, and the
maximum compliance limits are based on source range. See
more information.
NOTE The only exception to the compliance limit not being exceeded is the
2600S-901-01 Rev. C / January 2008
Model 2611/2612
Range
Source
200mV
±202mV
2V
±2.02V
20V
±20.2V
1
200V
±202V
100nA
±101nA
1µA
±1.01µA
10µA
±10.1µA
100µA
±101µA
1mA
±1.01mA
10mA
±10.1mA
100mA
±101mA
1A
±1.01A
1.5A
±1.515A
2
10A
±10.1A
Max Power = 30.603W per channel
1. 200V source range available only
when interlock is enabled. See
Section
10.
2. 10A range available only in pulse
mode.
VLIMIT when operating as an ISOURCE. To avoid excessive (and
potentially destructive) currents from flowing, the VLIMIT will source
or sink up to 102mA for ISOURCE ranges on or below 100mA. For
the ranges 1A and above, the maximum current allowed is the
current source setting.
Return to
Section Topics
Model 2635/2636
Measure
Range
±204mV
200mV
±2.04V
2V
±20.4V
20V
3
±204V
200V
±102nA
100pA
±1.02µA
1nA
±10.2µA
10nA
±102µA
100nA
±1.02mA
1µA
±10.2mA
10µA
±102mA
100µA
±1.02A
1mA
±1.53A
10mA
±10.2A
100mA
1A
1.5A
Max Power = 30.603W per channel
3. 200V source range available only
when interlock is enabled. See
Section
10.
"Compliance limit"
Section 4: Basic Operation
Source
Measure
+/-202mV
+/-204mV
+/-2.02V
+/-2.04V
+/-20.2V
+/-20.4V
+/-202V
+/-204V
N/A
+/-102pA
+/-1.01nA
+/-1.02nA
+/-10.1nA
+/-10.2nA
+/-101nA
+/-102nA
±1.01µA
±1.02µA
±10.1µA
±10.2µA
±101µA
±102µA
±1.01mA
±1.02mA
±10.1mA
±10.2mA
±101mA
±102mA
±1.01A
±1.02A
±1.515A
±1.53A
in Section 8 for
4-3