Series 2600 System SourceMeter® Instruments Reference Manual
WARNING

Calibration cycle

Perform calibration at least once a year to ensure the unit meets or exceeds its specifications.
Recommended calibration equipment
Table 16-1
equipment as long as that equipment has specifications equal to or greater than those listed in the
table. When possible, test equipment specifications should be at least four times better than
corresponding Series 2600 specifications.
2600S-901-01 Rev. C / January 2008
The maximum common-mode voltage (voltage between LO and
chassis ground) is 250VDC. Exceeding this value may cause a
breakdown in insulation, creating a shock hazard that could result
in personal injury or death.
The Input/Output terminals of the SourceMeters are rated for
connection to circuits rated Installation Category I only, with
transients rated less than 1500V peak. Do not connect the Series
2600 terminals to CAT II, CAT III, or CAT IV circuits. Connection of the
SourceMeter terminals to circuits higher than CAT I can cause
damage to the equipment or expose the operator to hazardous
voltage.
Hazardous voltages may be present on the output and guard
terminals. To prevent electrical shock that could cause injury or
death, NEVER make or break connections to the Series 2600 while
the unit is on. Power off the equipment from the front panel or
disconnect the main power cord from the rear of the Series 2600
before handling cables connected to the outputs. Putting the
equipment into standby mode does not guarantee the outputs are
not powered if a hardware or software fault occurs.
lists the recommended equipment for the calibration procedures. You can use alternate