Keithley 2601 Reference Manual page 200

System sourcemeter 2600 series (smu)
Hide thumbs Also See for 2601:
Table of Contents

Advertisement

Section 8: Source-Measure Concepts
Model 2611/2612/2635/2636 V-Source operating boundaries
Figure 8-9
covered. Operation in the other three quadrants is similar.
Figure
8-9A shows the output characteristics for the V-Source. As shown, the Model 2611/2612/
2635/2636 SourceMeter can output up to 20.2V at 1.5A, or 202V at 100mA. Note that when
sourcing more than 20.2V, current is limited to 100mA.
Figure
8-9B shows the limit lines for the V-Source. The voltage source limit line represents the
maximum source value possible for the presently selected voltage source range. For example, if
you are using the 20V source range, the voltage source limit line is at 20.2V. The current
compliance limit line represents the actual compliance in effect. See
this section. These limit lines are boundaries that represent the operating limits of the SourceMeter
for this quadrant of operation. The operating point can be anywhere inside (or on) these limit lines.
The limit line boundaries for the other quadrants are similar.
Figure 8-9
Model 2611/2612/2635/2636 V-Source boundaries
8-14
shows the operating boundaries for the V-Source. Only the first quadrant of operation is
Return to
Section Topics
Series 2600 System SourceMeter® Instruments Reference Manual
""Compliance
limit"" earlier in
2600S-901-01 Rev. C / January 2008

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

26362602261126122635

Table of Contents