Series 2600 System SourceMeter® Instruments Reference Manual
Remarks
• This attribute controls the SMU action taken when the output enable or interlock line is
asserted/de-asserted. The default setting is smuX.OE_NONE.
• This attribute is always set to OE_OUTPUT_OFF for the Model 2611/2612 and is read-
only.
When set to
•
interlock line goes low (de-asserted).
When set to
•
asserted, the SMU will turn its output off as if the
smuX.OUTPUT_OFF
command had been received.
The SMU will not automatically turn its output on when the output enable or interlock
•
line returns to the high state.
• If the output enable or interlock line is not asserted when this attribute is set to
smuX.OE_OUTPUT_OFF and the output is on, the output will turn off immediately.
• Detection of the output enable or interlock line going low will not abort any running
scripts. This may cause execution errors.
Details
See
"Output Enable (Models
smuX.source.offmode,
Also see
Example
Reconfigures the SMU to turn the output off if the output enable or interlock line goes low
(de-asserted):
smua.source.outputenableaction = smua.OE_OUTPUT_OFF
2600S-901-01 Rev. C / January 2008
, the SMU will take no action when the output enable or
smuX.OE_NONE
and the output enable or interlock line is de-
smuX.OE_OUTPUT_OFF
2601/2602)" in
smuX.source.output
Return to
Section Topics
smuX.source.output =
Section
10.
Section 12: Instrument Control Library
12-75