Section 12: Instrument Control Library
smuX.cal.restore
Function
Loads a stored set of calibration constants.
There are two ways to use this function:
Usage
smuX.cal.restore()
smuX.cal.restore(calset)
calset
Set calset
smuX.CALSET_NOMINAL
smuX.CALSET_FACTORY
smuX.CALSET_DEFAULT
smuX.CALSET_PREVIOUS
If
calset
Remarks
• This function will overwrite the current set of calibration constants from non-volatile
memory.
This function will be disabled until a successful call to
made.
Details
See
Section 16
Example
Restores factory calibration for SMU A:
smua.cal.restore(smua.CALSET_FACTORY)
smuX.cal.save
Function
Stores the calibration constants in non-volatile memory.
Usage
smuX.cal.save()
Remarks
• This function will store the current set of calibration constants in non-volatile memory. The
previous calibration constants (from the default calibration set) will be copied to the
previous calibration set (smuX.CALSET_PREVIOUS) prior to overwriting the default
calibration set.
This function will be disabled until a successful call to
made.
been changed, this function will be disabled unless both the calibration date and the
calibration due date have been assigned new values. For models 2635 and 2636: If any
of the calibration constants have been changed, this function will be disabled, unless the
calibration adjust date has been set.
Details
See
Section 16
Also see
smuX.cal.date, smuX.cal.due,
Example
Stores calibration constants for SMU A in non-volatile memory:
smua.cal.save()
smuX.cal.state
Attribute
Calibration state.
Usage
calstate = smuX.cal.state
12-52
X = SMU channel (a or b)
to one of the following values:
is not specified,
smuX.CALSET_DEFAULT
"Calibration".
X = SMU channel (a or b)
For models 2601, 2602, 2611 and 2612: If any of the calibration constants have
"Calibration".
smuX.cal.restore
X = SMU channel (a or b)
Return to
Series 2600 System SourceMeter® Instruments Reference Manual
Calibration set to be loaded.
A set of calibration constants that are
uncalibrated but set to nominal values to
allow rudimentary functioning of the
instrument.
The calibration constants when the instrument
left the factory.
The normal calibration set.
The calibration set that was used before the
last
default set was overwritten.
will be used.
smuX.cal.unlock
smuX.cal.unlock
Section Topics
is
is
2600S-901-01 Rev. C / January 2008