Section 9: System Expansion (TSP-Link)

Resetting the TSP-Link

After all the node numbers are set, you must initialize the system by performing a TSP-Link reset.
For initialization to succeed, all units must be powered on when the TSP-Link reset is performed.
NOTE If you change the system topology after initialization, you must
Front panel operation
Complete the following steps to reset the TSP-Link network from the front panel.
1.
Press Menu > TSPLINK.
2.
Press RESET.
Remote programming
The commands associated with TSP-Link reset are listed in
Table 9-1
TSP-Link reset commands
Command
tsplink.reset()
tsplink.state
An attempted TSP-Link reset will fail if any of the following conditions are true:
Two or more instruments in the system have the same node number.
There are no other instruments connected to the unit performing the reset.
One or more of the units in the system is not powered on.
Programming example – The following code will reset the TSP-Link and output its state:
tsplink.reset()
print(tsplink.state)
If the reset is successful,
have been established.
9-4
re-initialize the system by performing a TSP-Link reset. Changes that
affect the system topology include:
Powering down or rebooting any unit in the system.
Rearranging or disconnecting the LAN cable connections between
units.
Description
Initializes the TSP-Link network.
Returns online if the most recent TSP-Link reset was
successful. Returns offline if the reset failed.
will be returned to indicate that communications with all nodes
online