Section 13: Factory Scripts
Table 13-4 (continued)
KIGeneral TSP test script: SweepVLinMeasureI
TSP project name:
TSP test script name:
Firmware version:
Parameters
Data
Example
13-6
KIFactoryGeneral
KIGeneral
1.0.2 and later
smu, startv, stopv, stime, points
SourceMeter Channel (A or B). Defaults to SMU A if all
smu:
parameters are omitted when function is called.
Sweep start voltage in volts.
starti:
Sweep stop voltage in volts.
stopi:
Settling time in seconds. Occurs after stepping the source
stime:
and before performing a measurement.
Number of sweep points (must be
points:
Current measurements, voltage source values and timestamps are stored in
smuX.nvbuffer1.
SweepVLinMeasureI(smua, -1, 1, 1E-3, 1000)
This function performs a 1000-point linear voltage sweep starting at -1V and
stopping at +1V. Current is measured at every step (point) in the sweep after a
1ms source settling period.