Section 16: Calibration
Models 2635 and 2636:
1.
Connect the Series 2600 to the digital multimeter (see
multimeter DC current function.
2.
Calibrate the low current ranges (100pA, 1nA, 10nA, 100nA
characterized 1G
RES. (see
connection diagrams, and a factory script for calibrating the low current ranges are included
with the Model 2600-STD-RES. The general process entails forcing a characterized voltage
across the 1G
standard resistance and voltage derived current.
a.
b.
Table 16-6
Settings of Model 2635/2636 Characterization of Voltage Source
Low Current Range
100pA
1nA
10nA
100nA
Step 4. Contact check calibration (For 2601/2602/2611/2612 only)
a. Short the Series 2600 sense low and output low terminals, as shown in
1.The 2601/02/11/12 could be calibrated with this method for the 100nA setting only if desired.
16-18
Ω
resistance standard, such as the Keithley Instruments Model 2600-STD-
Table
16-1,
"Recommended calibration equipment"
Ω
resistor and comparing the 2635/2636 measured results against the
Characterize the appropriate +/- V source values with the Digital Multimeter according
to
Table
16-4.
Characterize the desired Model 2635/2636 current ranges.
1) Connect the guarded resistance standard.
2) Source the appropriate voltage for +/- full scale reading.
3) Wait 30 seconds for stable measurement.
4) Capture the Model 2635/2636 reported current measurement.
5) Initiate HI-Z mode to open the resistor standard (source zero current) and the
characterize offset.
6) Repeat the above steps for each low current range.
Voltage Source
+/- 100.00mV
+/- 1.0000mV
+/- 10.000mV
+/- 100.00V
Also short the sense high and output high terminals together, as shown in the figure.
Return to
Section Topics
Series 2600 System SourceMeter® Instruments Reference Manual
Figure
Compliance
100mA
16-2), and select the
1
) using a suitably guarded and
).Step-by-step procedures,
1.5A
1.5A
1.5A
Figure
16-4.
2600S-901-01 Rev. C / January 2008