Input/Output Lo And Chassis Ground - Keithley 2601 Reference Manual

System sourcemeter 2600 series (smu)
Hide thumbs Also See for 2601:
Table of Contents

Advertisement

Series 2600 System SourceMeter® Instruments Reference Manual
Figure 3-1
2602/2612 input/output connectors
S
G
HI
Figure 3-2
Model 2636 input/output connectors

Input/output LO and chassis ground

As shown in
Input/output LOs are not connected between channels and are electrically isolated from chassis
ground.
2600S-901-01 Rev. C / January 2008
Channel B
G
G
HI
G
S
LO
LO
CHANNEL B
HI = Input/Output HI
S HI = Sense HI
G = Guard
S LO = Sense LO
LO = Input/Output LO
Figure
3-3, SMU input/output LOs are available at the rear panel terminal blocks.
Return to
Channel A
CHANNEL A
S
LO
LO
HI
G
Captive screw (2 per terminal block)
Each terminal block uses two captive
screws to secure it to the rear panel.
Section Topics
Section 3: DUT Test Connections
S
G
G
G
HI
3-3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

26362602261126122635

Table of Contents