Section 12: Instrument Control Library
Remarks
This function will write the given string to the serial port where it can be read by equipment
(e.g., component handler) connected to the other end of the serial port. No terminator
characters are added to the data. The
Also see
serial.read
Writes data string "1 2 3 4" to the serial port:
Example
serial.write("1 2 3 4")

setup functions and attribute

The functions and attribute in this group are used to save/recall setups and to set the power-on setup.
setup.poweron
Attribute
The saved setup to recall when the unit is turned on.
Usage
n = setup.poweron
setup.poweron = n
n
Remarks
For an n setting of 0, the unit powers up to the factory default (reset) setup. For an
n setting of 1 to 5, the unit powers up to a user saved setup.
Details
See
"Remote operation
Example
Sets unit to power on to the factory defaults:
setup.poweron = 0
setup.recall
Function
Recalls settings from a saved setup.
Usage
setup.recall(n)
n
For an
Remarks
n
1 to 5, the unit recalls a user saved setup.
of
Details
See
"Remote operation
Example
Recalls the user-setup at location 2:
setup.recall(2)
setup.save
Function
Saves the present setup as a user-setup.
Usage
setup.save(n)
n
Numbers 1 through 5 are used to designate user-setup locations. When you save
Remarks
at one of these locations, the previous setup at that location is overwritten.
Details
See
"Remote operation
Example
Saves the present setup at location 5:
setup.save(5)
smuX functions and attributes
The functions and attributes in this group are used to control basic source-measure operations of
the SMUs and perform calibration.
smuX.cal.adjustdate
Attribute
Adjustment date of the last calibration adjustment
12-48
setups" in
Section
setting of 0, the unit recalls the factory default (reset) setup. For an
setups" in
Section
setups" in
Section
X= SMU channel (a or b)
(Models 2635 and 2636 only)
Return to
Series 2600 System SourceMeter® Instruments Reference Manual
data
will be written exactly as is.
-- Reads the power-on setup.
-- Writes the power-on setup.
-- Setup number to recall on power up (0 to 5).
1.
Setup number to recall (0 to 5).
1.
Setup number to save (1 to 5).
1.
Section Topics