Current Programming Accuracy
Range
100.000nA
1.00000 A
10.0000 A
100.000 A
1.00000mA
10.0000mA
100.000mA
2
1.00000A
2
1.50000A
2,5
10.0000A
Temperature coefficient (0°C–18°C and 28°C–50°C): ±(0.15 × accuracy specification)/°C.
Maximum output power and source/sink limits:
at ±202V, four-quadrant source or sink operation.
Current regulation: Line: 0.01% of range. Load: ±(0.01% of range + 100pA).
Voltage limit/compliance:
Accuracy same as voltage source.
Overshoot: <0.1% typical (step size = 10% to 90% of range, resistive load; see "Current source output settling
time" for additional test conditions).
Additional Source Specifications
Transient response time: <70 s for the output to recover to 0.1% for a 10% to 90% step change in load.
Voltage source output settling time: Time required to reach 0.1% of final value after source level command is processed on a
fixed range.
200mV, 2V ranges: <50 s typical. 20V range: <110 s typical. 200V range: <700 s typical.
Current source output settling time: Time required to reach 0.1% of final value after source level command is processed on a
fixed range. Values below for Iout · Rload = 2V unless noted.
1.5A–1A ranges: <120 s typical (Rload >6 ).
100mA–10mA ranges: <80 s typical.
1mA range: <100 s typical.
100 A range: <150 s typical.
4
Accuracy specifications do not include connector leakage. Derate accuracy by Vout/2E11 per °C when operating
between 18°C–28°C. Derate accuracy by Vout/2E11 + (0.15 * Vout/2E11) per °C when operating <18°C and >28°C.
5
10A range accessible only in pulse mode.
6
For sink mode operation (quadrants II and IV), add 10% of compliance range and ±0.02% of limit setting to corresponding
voltage source specification. For 200mV range add an additional 120mV of uncertainty.
Specifications are subject to change without notice.
Page 2 of 6
4
Programming
resolution
2pA
20pA
200pA
2nA
20nA
200nA
2 A
20 A
50 A
200 A
2
6
Bipolar voltage limit (compliance) set with a single value. Minimum value is 10mV.
System SourceMeter
Accuracy (1 year)
23°C + 5°C
+ (% rdg.+amps)
0.06% + 100pA
0.03% + 800pA
0.03% +
5nA
0.03% + 60nA
0.03% + 300nA
0.03% +
6 A
0.03% + 30 A
0.05% + 1.8mA
0.06% + 4mA
0.5% + 40mA
30.603W per channel maximum. ±1.515A at ±20.2V, ±101mA
Models 2611/2612
®
Specifications
Typical noise
(peak-peak)
0.1Hz-10Hz
5pA
25pA
60pA
3nA
6nA
200nA
600nA
70 A
150 A
SPEC-2611_2612 Rev. A / June 2006