Series 2600 System SourceMeter® Instruments Reference Manual
4-wire remote sensing
When sourcing and/or measuring voltage in a low-impedance test circuit (see
can be errors associated with IR drops in the test leads. Voltage source and measure accuracy are
optimized by using 4-wire remote sense connections. When sourcing voltage, 4-wire remote
sensing ensures that the programmed voltage is delivered to the DUT. When measuring voltage,
only the voltage drop across the DUT is measured.
Use 4-wire remote sensing for the following source-measure conditions:
•
Sourcing and/or measuring voltage in low impedance (<1kΩ) test circuits.
•
Enforce voltage compliance limit directly at the DUT.
Figure 3-6
Model 2602/2612 four-wire connections
Model 2636 four-wire connections (remote sensing)
2600S-901-01 Rev. C / January 2008
(remote sensing)
Return to
Section Topics
Section 3: DUT Test Connections