Table 13-11 (continued)
KIGeneral TSP test script: ConfigPulseVMeasureI
TSP project name:
TSP test script name:
Firmware version:
Function:
ConfigPulseVMeasureI
Usage:
f, msg = ConfigPulseVMeasureI
(smu,bias,level,limit,ton,toff,points,buffer,tag[,sync_in][,sync_out]
[,sync_in_timeout][,sync_in_abort])
Waveform
Return
Values
Output
Data
Example
See Also
2600S-901-01 Rev. C / January 2008
KIFactoryPulse
KIPulse
1.2.0 and later
A Boolean flag. This flag will be true when the pulse is successfully
f:
configured, false when errors were encountered.
msg: A string message. If the f flag is false, msg will contain an error
message. Otherwise, msg will contain a string indicating successful
configuration.
Pulsed current measurements are stored in the reading buffer specified by the
buffer input argument.
ConfigPulseVMeasureI(smub, 0, 20, 1, 0.001, 0.080, 10,
smub.nvbuffer1, 2)
Set up a pulse train that will use SourceMeter channel B. The pulse amplitude will
be 20V and will return to 0V after 1msec. The pulse will remain at 0V for 80msec
and the current limit will be 1A during the pulse. The pulse train will consist of 10
pulses and the pulse train will be assigned a tag index of 2.
InitiatePulseTest,
InitiatePulseTestDual
Return to
Section Topics
Section 13: Factory Scripts
13-17