Section 12: Instrument Control Library
Include source values with readings for SMU A buffer 1:
Example
smua.nvbuffer1.collectsourcevalues = 1
smuX.nvbufferY.collecttimestamps
Attribute
Timestamp collection for the buffer.
Usage
state = smuX.nvbufferY.collecttimestamps
smuX.nvbufferY.collecttimestamps = state
Set
state
0 Timestamp collection disabled (off).
1 Timestamp collection enabled (on).
Remarks
• Assigning a state value to this attribute enables or disables the storage of timestamps.
Reading this attribute returns the state of timestamp collection.
• When on, timestamps will be stored with readings in the buffer. This requires four extra
bytes of storage per reading. The first reading is time stamped at zero seconds.
Subsequent readings are time stamped relative to the time storage was started.
• This value, off or on, can only be changed when the buffer is empty. The buffer can be
emptied using the
Details
See
"Reading
Also see
smuX.measure.overlappedY, smuX.measure.Y,
Include timestamps with readings for SMU A buffer 1:
Example
smua.nvbuffer1.collecttimestamps = 1
smuX.nvbufferY.n
Attribute
Number of readings in the buffer.
Usage
bufferreadings = smuX.nvbufferY.n
Remarks
• Reading this attribute returns the number of readings that are stored in the buffer.
• This is a read-only attribute.
Details
See
"Reading
Also see
smuX.measure.overlappedY, smuX.measure.Y,
Example
Read the number of readings stored in SMU A buffer 1:
bufferreadings = smua.nvbuffer1.n
print(bufferreadings)
Output:
1.250000+02
The above output indicates that there are 125 readings stored in the buffer.
smuX.nvbufferY.timestampresolution
Attribute
Timestamp resolution.
Usage
tsres = smuX.nvbufferY.timestampresolution
smuX.nvbufferY.timestampresolution = tsres
tsres
Remarks
• Assigning to this attribute sets the resolution for the timestamps. Reading this attribute
returns the timestamp resolution value.
• The minimum timestamp resolution is 0.000001seconds (1µs). At this resolution, the
reading buffer can store unique timestamps for up to 71 minutes. This value can be
increased for very long tests.
• When setting this value it will be rounded to an even power of 2µs.
Details
See
"Reading
Also see
smuX.measure.overlappedY, smuX.measure.Y,
Set the timestamp resolution for SMU A buffer 1 to 10µs:
Example
smua.nvbuffer1.timestampresolution = 0.00001
12-68
X = SMU channel (a or b)
Y = NV buffer (1 or 2)
to one of the following values:
smuX.nvbufferY.clear
buffers" in this section and in
X = SMU channel (a or b)
Y = NV buffer (1 or 2)
buffers" in this section and in
X = SMU channel (a or b)
Y = NV buffer (1 or 2)
Timestamp resolution in seconds.
buffers" in this section and in
Return to
Section Topics
Series 2600 System SourceMeter® Instruments Reference Manual
-- Reads collection state.
-- Writes collection state.
function.
Section
7.
smuX.nvbufferY
Section
7.
smuX.nvbufferY
-- Reads collection state.
-- Writes collection state.
Section
7.
smuX.nvbufferY