Section 12: Instrument Control Library
status.measurement.*
status.measurement.condition
status.measurement.enable
status.measurement.event
status.measurement.ntr
status.measurement.ptr
* = condition, enable, event, ntr or ptr
Attribute
Measurement event status register set.
Reads condition, enable, event, NTR and PTR registers:
Usage
measreg = status.measurement.condition
measreg = status.measurement.current_limit.enable
measreg = status.measurement.current_limit.event
measreg = status.measurement.current_limit.ntr
measreg = status.measurement.current_limit.ptr
Writes to enable, NTR and PTR registers:
status.measurement.enable = measreg
status.measurement.ntr = measreg
status.measurement.ptr = measreg
Set measreg to one of the following values:
0
status.measurement.VOLTAGE_LIMIT
status.measurement.VLMT
status.measurement.CURRENT_LIMIT
status.measurement.ILMT
status.measurement.READING_OVERFLOW
status.measurement.ROF
status.measurement.BUFFER_AVAILABLE
status.measurement.BAV
status.measurement.OUTPUT_ENABLE
status.measurement.OE
status.measurement.INSTRUMENT_SUMMARY
status.measurement.INST
measreg
To set bit B0 (VLMT), set
To set bit B1 (ILMT), set
To set bit B8 (BAV), set
To set more than one bit of the register, set measreg to the sum of their decimal weights.
For example, to set bits B0 and B8, set measreg to 257 (1 + 256).
Remarks
• This attribute is used to read or write to the measurement registers.
• Reading a status register returns a value. The binary equivalent of the returned value
indicates which register bits are set. The least significant bit of the binary number is bit 0,
and the most significant bit is bit 15.
• For example, assume value 257 is returned for the register. The binary equivalent is
0000000100000001. This value indicates that bit B1(VLMT) and bit B8 (BAV) are set.
• The used bits of the measurement registers are described as follows:
Bit B0, VLMT – Set bit indicates that the voltage limit was exceeded. This bit will
be updated only when a measurement is taken or
invoked.
Bit B1, ILMT – Set bit indicates that the current limit was exceeded. This bit will be
updated only when a measurement is taken or
invoked.
Bit B7, ROF –
Bit B8, BAV –
the non-volatile reading buffers.
Bit B11, OE –
Bit B13, INST –
register is set.
Details
See
"Measurement Event
12-78
can also be set to the decimal weight of the bit to be set. Examples:
measreg
to 2 (2
measreg
to 256 (2
measreg
Set bit indicates that an overflow reading has been detected.
Set bit indicates that there is at least one reading stored in either or both of
Set bit indicates that output enable has been asserted.
Set bit indicates that a bit in the measurement instrument summary
Registers" in
Return to
Section Topics
Series 2600 System SourceMeter® Instruments Reference Manual
Clears all bits.
Sets VLMT bit (B0).
Sets VLMT bit (B0).
Sets ILMT bit (B1).
Sets ILMT bit (B1).
Sets ROF bit (B7).
Sets ROF bit (B7).
Sets BAV bit (B8).
Sets BAV bit (B8).
Sets OE bit (B11).
Sets OE bit (B11).
Sets INST bit (B13).
Sets INST bit (B13).
0
to 1 (2
).
1
).
8
).
smuX.source.compliance
smuX.source.compliance
Appendix
D.
is
is
2600S-901-01 Rev. C / January 2008