Series 2600 System SourceMeter® Instruments Reference Manual
Table 13-16 (continued)
KIGeneral TSP test script: ConfigPulseVMeasureISweepLog
TSP project name:
TSP test script name:
Firmware version:
Function:
ConfigPulseVMeasureISweepLog
Usage:
f, msg = ConfigPulseVMeasureISweepLog
(smu,bias,start,stop,limit,ton,toff,points,buffer,tag[,sync_in]
[,sync_out]
[,sync_in_timeout][,sync_in_abort]
Configuration
Return
Values
Output
Data
Example
See Also
2600S-901-01 Rev. C / January 2008
KIFactoryPulse
KIPulse
1.2.0 and later
A Boolean flag. This flag will be true when the pulse was successfully
f:
configured, false when errors were encountered.
A string message. If the f flag is false, msg will contain an error
msg:
message. Otherwise, msg will contain a string indicating successful
configuration.
Pulsed current measurements are stored in the reading buffer specified by the
buffer input argument.
ConfigPulseVMeasureISweepLog(smub, 0, 1, 10, 1, 10e-3,
20e-3, 10, smub.nvbuffer1, 6)
Set up a pulsed log sweep that will use SourceMeter channel B. The pulsed
sweep will start at 1V, end at 10V, and return to a 0V bias level between pulses.
Each pulsed-step will be on for 10msec and then at the bias level for 20msec. The
current limit will be 1A during the entire pulsed-sweep. The pulse train will be
comprised of 10 pulsed-steps and the pulse train will be assigned a tag index of 6.
InitiatePulseTest,
InitiatePulseTestDual
Return to
Section Topics
)
Section 13: Factory Scripts
13-27