Series 2600 System SourceMeter® Instruments Reference Manual
Remote rel programming
Rel commands
Rel commands are summarized in
Table 6-5
Rel commands
Command
To set rel values:
smuX.measure.rel.leveli = relval
smuX.measure.rel.levelp = relval
smuX.measure.rel.levelr = relval
smuX.measure.rel.levelv = relval
To enable/disable rel:
smuX.measure.rel.enablei = smuX.REL_OFF
smuX.measure.rel.enablep = smuX.REL_OFF
smuX.measure.rel.enabler = smuX.REL_OFF
smuX.measure.rel.enablev = smuX.REL_OFF
smuX.measure.rel.enablei = smuX.REL_ON
smuX.measure.rel.enablep = smuX.REL_ON
smuX.measure.rel.enabler = smuX.REL_ON
smuX.measure.rel.enablev = smuX.REL_ON
* smuX = smua for the Model 2601/2611; smuX = smua (Channel A) or smub (Channel B) for the
Model 2602/2612.
Rel programming example
2600S-901-01 Rev. C / January 2008
1
smua.reset()
--Restore Series 2600 defaults.
smua.measure.rel.leveli = 0.1
--Set current rel to 100mA.
smua.measure.rel.enablei = smua.REL_ON
--Enable current rel.
Return to
Section Topics
Section 6: Range, Digits, Speed, Rel, and Filters
Table
6-5.
Description
Set current rel value.
Set power rel value.
Set resistance rel value.
Set voltage rel value.
Disable current rel.
Disable power rel.
Disable resistance rel.
Disable voltage rel.
Enable current rel.
Enable power rel.
Enable resistance rel.
Enable voltage rel.
6-9