Section 13: Factory Scripts
Table 13-5 (continued)
KIGeneral TSP test script: SweepILogMeasureV
TSP project name:
TSP test script name:
Firmware version:
Function:
SweepILogMeasureV(smu, starti, stopi, stime, points)
Example
Table 13-6
KIGeneral TSP test script: SweepVLogMeasureI
TSP project name:
TSP test script name:
Firmware version:
Function:
SweepVLogMeasureI(smu, startv, stopv, stime, points)
13-8
KIFactoryGeneral
KIGeneral
1.0.2 and later
SweepILogMeasureV(smua, 0.01, 0.1, 0.001, 5)
This function performs a 5-point logarithmic current sweep starting at 10mA and
stopping at 100mA. Voltage is measured at every step (point) in the sweep. The
source will be allowed to settle on each step for 1ms before a measurement is
performed.
The following log values and corresponding source levels for the 5-point log
sweep are listed as follows:
MeasurePoint
LogStepSize
1
0.25
2
0.25
3
0.25
4
0.25
5
0.25
KIFactoryGeneral
KIGeneral
1.0.2 and later
Return to
Section Topics
Series 2600 System SourceMeter® Instruments Reference Manual
LogStep
SourceStepLevel
0.0
0.01A
0.25
0.017783A
0.5
0.031623A
0.75
0.056234A
1.0
0.1A