Keithley 2601 Reference Manual page 484

System sourcemeter 2600 series (smu)
Hide thumbs Also See for 2601:
Table of Contents

Advertisement

Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
www.keithley.com
Pulse width programming resolution: 1 s.
Pulse width programming accuracy:
Typical pulse width jitter: 50 s.
3. METER SPECIFICATIONS
Voltage Measurement Accuracy
Range
200.000mV
2.00000V
20.0000V
200.000V
Temperature coefficient (0°C–18°C and 28°C–50°C): ±(0.15 × accuracy specification)/°C.
11 Add 50 V to source accuracy specifications per volt of HI lead drop.
12 De-rate accuracy specifications for NPLC setting <1 by increasing error term. Add appropriate % of range term using
table below:
NPLC
setting
0.1
0.01
0.001
13 Applies when in single channel display mode.
Specifications are subject to change without notice.
Page 4 of 6
7
±25 s.
+10A
+1.5A
3
+1A
2
+0.1A
0A
–0.1A
2
–1A
3
–1.5A
–10A
–200V
–180V
–20V
11, 12
Display resolution
1 V
10 V
100 V
1mV
200mV
2V–200V
range
ranges
0.01%
0.01%
0.08%
0.07%
0.8 %
0.6 %
System SourceMeter
4
1
4
0V
+5V
+20V
–5V
13
Input resistance
>10G
>10G
>10G
>10G
100nA
1 A–100mA
range
0.01%
0.1 %
1 %
Models 2611/2612
®
Specifications
3
2
2
3
+180V
+200V
Accuracy (1 year)
23°C + 5°C
+ (% rdg.+ volts)
0.015% + 225 V
0.02% + 350 V
0.015% + 5mV
0.015% + 50mV
1A–1.5A
ranges
ranges
0.01%
0.01%
0.05%
0.05%
0.5 %
1.1 %
SPEC-2611_2612 Rev. A / June 2006

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

26362602261126122635

Table of Contents