Series 2600 System SourceMeter® Instruments Reference Manual
Table 6-2
Range commands
1
Commands
Measure range commands:
smuX.measure.autorangei = smuX.AUTORANGE_ON
smuX.measure.autorangei = smuX.AUTORANGE_OFF
smuX.measure.autorangev = smuX.AUTORANGE_ON
smuX.measure.autorangev = smuX.AUTORANGE_OFF
smuX.measure.lowrangei = lowrange
smuX.measure.lowrangev = lowrange
smuX.measure.rangei = rangeval
smuX.measure.rangev = rangeval
Source range commands:
smuX.source.autorangei = smuX.AUTORANGE_ON
smuX.source.autorangei = smuX.AUTORANGE_OFF
smuX.source.autorangev = smuX.AUTORANGE_ON
smuX.source.autorangev = smuX.AUTORANGE_OFF
smuX.source.limiti = level
smuX.source.limitv = level
smuX.source.lowrangei = lowrange
smuX.source.lowrangev = lowrange
smuX.source.rangei = rangeval
smuX.source.rangev = rangeval
* smuX = smua for the Model 2601/2611; smuX = smua (Channel A) or smub (Channel B) for the Model 2602/2612.
** See
Table 6-1
for measure ranges.
*** See Table 6-1 for source ranges
Range programming example
The listing below shows a programming example for controlling both source and measure ranges.
The SourceMeter is set up as follows:
Voltage source range: auto
Current measure range: 10mA
Measure low range: = 10μA
Voltage source current limit: 10mA
2600S-901-01 Rev. C / January 2008
2
2
smua.reset()
--Restore Series 2600 defaults.
smua.source.autorangev = smua.AUTORANGE_ON
--Set V source range to auto.
smua.measure.rangei = 1e-2
--Enable 10mA measure range.
smua.measure.lowrangei = 1e-5
--Set lowest range to 10μA.
smua.source.limiti = 1e-2
--Set limit level to 10mA.
Return to
Section Topics
Section 6: Range, Digits, Speed, Rel, and Filters
Description
Enable current measure auto range.
Disable current measure auto range.
Enable voltage measure auto range.
Disable voltage measure auto range.
Set lowest I measure range for auto range.
Set lowest V measure range for auto range.
Select manual current measure range.
Select manual voltage measure range.
Enable current source auto range.
Disable current source auto range.
Enable voltage source auto range.
Disable voltage source auto range.
Set voltage source current limit.
Set current source voltage limit.
Set lowest I source range for auto range.
Set lowest V source range for auto range.
Select manual current source range.
Select manual voltage source range.
6-5