Section 12: Instrument Control Library
Remarks
• Reading this attribute gives the output-off mode of the source. Setting this attribute
configures the SMU output-off mode.
• The default offmode is smuX.OUTPUT_NORMAL. In this mode, the SMU will source 0
volts. For the 2601/2602, the compliance to 10% of the current source range or 100uA,
whichever is smaller. If the source function is voltage, the 10% compliance will inherently
be a reduction in compliance current. If the source function is current, the 10%
compliance value may be more or less than the current that was being sourced. For the
2611/2612/2635/2636, the compliance to the value specified by
smuX.source.offlimiti (default 1mA).
• When
when the output is turned off.
• When the offmode is set to smuX.OUTPUT_ZERO, the SMU will source 0 volts just as
OUTPUT_NORMAL mode does. If the source function is voltage, the current limit will not be
changed. If the source function was current, the current limit will be set to the current
source level or 10% of the current source range, whichever is greater.
Details
See
"Output-off
Also see
smuX.source.output
Sets output-off mode for SMU A:
Example
smua.source.offmode = smua.OUTPUT_HIGH_Z
smuX.source.output
Attribute
Source output control (on/off).
Usage
state = smuX.source.output
smuX.source.output = state
Set
state
or
smuX.OUTPUT_OFF
0
or
smuX.OUTPUT_ON
1
Reading this attribute gives the output state of the source. Setting this attribute will turn
Remarks
the output of the source on or off. The default for the source is off. When the output is
turned on, the SMU will source either voltage or current as dictated by the
smuX.source.func
Details
See
Section 4
Also see
smuX.source.func,
Turns SMU A source output on:
Example
smua.source.output = smua.OUTPUT_ON
smuX.source.outputenableaction
Attribute
Output enable or interlock action for the source.
Usage
action = smuX.source.outputenableaction
smuX.source.outputenableaction = action
Set
action
or
smuX.OE_NONE
0
or
smuX.OE_OUTPUT_OFF
1
12-74
offmode
is set to smuX.OUTPUT_HIGH_Z, the SMU will open the output relay
.
states" in
Section 3
X = SMU channel (a or b)
to one of the following values:
setting.
"basic operation".
smuX.source.offmode
X = SMU channel (a or b)
to one of the following values:
Return to
Series 2600 System SourceMeter® Instruments Reference Manual
-- Reads output state.
-- Writes output state.
Turns the source output off.
Turns the source output on.
-- Reads enable action.
-- Writes enable action.
No action.
Turns the source output off.
Section Topics