Series 2600 System SourceMeter® Instruments Reference Manual
Similarly, the following would return 100 Channel A source values from buffer 1:
printbuffer(1, 100, smua.nvbuffer1.sourcevalues)
Note that readings is the default reading attribute and can be omitted. Thus, the following would
also return 100 Channel A readings from buffer 1:
printbuffer(1, 100, smua.nvbuffer1)
Table 7-6
Recall attributes
1
Recall attribute
measurefunctions
measureranges
readings
sourcefunctions
sourceoutputstates
sourceranges
sourcevalues
statuses
timestamps
1. The default attribute is readings and can be omitted. For example, smua.nvbuffer1 and
smua.nvbuffer1.readings will both return readings from Channel A, buffer 1.

Time and date values

All time and date values are represented as the number of seconds since the unit was powered
on. The os.clock() function returns values in this format.
Representing time as the number of seconds will be referred to as "standard time format." Note
that because TSL numbers are floating point numbers, the precision of time and date values
relative to power-on will decrease the longer the unit is powered on. This decrease in precision is
approximately 0.06ppm of the total elapsed time. This precision is generally much better than the
time base of the instrument and should not present any problems in practice. It is worth noting
because the user can directly see the affects as compared to the less obvious time-base drift.
2600S-901-01 Rev. C / January 2008
Description
An array (a LUA table) of strings indicating the function measured for
the reading (Current, Voltage, Ohms or Watts).
An array (a LUA table) of full scale range values for the measure range
used when the measurement was made.
An array (a LUA table) of the readings stored in the reading buffer. This
array holds the same data that is returned when the reading buffer is
accessed directly, i.e., rb[2] and rb.readings[2] are the same value.
An array (a LUA table) of strings indicating the source function at the
time of the measurement (Current or Voltage).
An array (a LUA table) of strings indicating the state of the source (Off
or On).
An array (a LUA table) of full scale range values for the source range
used when the measurement was made.
If enabled, an array (a LUA table) of the sourced values in effect at the
time of the reading.
An array (a LUA table) of status values for all of the readings in the
buffer. The status values are floating-point numbers that encode the
status value into a floating-point value (see
An array (a LUA table) of time stamps, in seconds, of when each read-
ing occurred. These are relative to the basetime stamp for the buffer
(Table
7-3).
Return to
Section Topics
Section 7: Buffer (Data Store)
Table
7-7).