Section 3: DUT Test Connections
NOTE The OUTPUT OFF STATE menu can also be accessed by navigating
With the OUTPUT OFF STATE menu displayed, select the desired output-off state: HI-Z
(high-impedance), NORMAL, or ZERO.
Remote programming
Table 3-3
Table 3-3
Commands to select output-off state
Command
smuX.source.offlimiti = ivalue
smuX.source.offmode = smuX.OUTPUT_NORMAL
smuX.source.offmode = smuX.OUTPUT_HIGH_Z
smuX.source.offmode = smuX.OUTPUT_ZERO
* Model 2601/2611/2635: smuX = smua, Model 2602/2612/2636: smuX = smua (Channel A) or
smub (Channel B).
3-24
the configuration menu that is displayed by pressing the CONFIG
key.
lists the commands to select the output-off state.
*
Return to
Series 2600 System SourceMeter® Instruments Reference Manual
Description
Sets current limit in normal output-
off state for firmware revision
1.2.0, or later.
Selects normal output-off state.
Selects high-impedance output-
off state.
Selects zero output-off state.
Section Topics
2600S-901-01 Rev. C / January 2008