Using Shielding And Guarding Together - Keithley 2601 Reference Manual

System sourcemeter 2600 series (smu)
Hide thumbs Also See for 2601:
Table of Contents

Advertisement

Section 3: DUT Test Connections
Figure 3-20
Model 2636 connections for test circuit shown in

Using shielding and guarding together

Figure 3-21
guarding. The guard shields are connected to the driven guard (G) of the SMU. The noise shield is
connected to SMU LO. The safety shield is connected to the chassis and to a safety earth ground.
Figure 3-21
Model 2601/2602-1 connections for noise shield, safety shield, and guarding
3-18
shows connections for a test system that uses a noise shield, a safety shield, and
Return to
Section Topics
Series 2600 System SourceMeter® Instruments Reference Manual
Figure 3-18
2600S-901-01 Rev. C / January 2008