Appendix D: Status Model

Service request enable register

The generation of a service request is controlled by the Service Request Enable Register. This
register is programmed by the user and is used to enable or disable the setting of bit B6 (RQS/
MSS) by the Status Summary Message bits (B0, B2, B3, B4, B5, and B7) of the Status Byte
Register. As shown in
corresponding enable bits of the Service Request Enable Register. When a set (1) summary bit is
ANDed with an enabled (1) bit of the enable register, the logic "1" output is applied to the input of
the OR gate and, therefore, sets the MSS/RQS bit in the Status Byte Register.
The individual bits of the Service Request Enable Register can be set or cleared by using the
*SRE common command or its script equivalent. To read the Service Request Enable Register,
use the *SRE? query or script equivalent. The Service Request Enable Register clears when
power is cycled or a parameter value of 0 is sent with the *SRE command (i.e. *SRE 0). The
commands to program and read the SRQ Enable Register are listed in
Serial polling and SRQ
Any enabled event summary bit that goes from 0 to 1 will set bit B6 and generate an SRQ (service
request). In your test program, you can periodically read the Status Byte to check if an SRQ has
occurred and what caused it. If an SRQ occurs, the program can, for example, branch to an
appropriate subroutine that will service the request.
SRQs can be managed by the serial poll sequence of the SourceMeter. If an SRQ does not occur,
bit B6 (RQS) of the Status Byte Register will remain cleared, and the program will simply proceed
normally after the serial poll is performed. If an SRQ does occur, bit B6 of the Status Byte Register
will set, and the program can branch to a service subroutine when the SRQ is detected by the
serial poll.
The serial poll automatically resets RQS of the Status Byte Register. This allows subsequent serial
polls to monitor bit B6 for an SRQ occurrence generated by other event types.
For common and script commands, B6 is the MSS (Message Summary Status) bit. The serial poll
does not clear MSS. The MSS bit stays set until all Status Byte summary bits are reset.
SPE, SPD (serial polling)
For the GPIB interface only, the SPE, SPD General Bus Command sequence is used to serial poll
the SourceMeter (see
byte (status byte). Typically, serial polling is used by the controller to determine which of several
instruments has requested service with the SRQ line.
Status byte and service request commands
The commands to program and read the Status Byte Register and Service Request Enable
Register are listed in
script command equivalents. For details on programming and reading registers, see
"Programming enable and transition registers"
To reset the bits of the Service Request Enable Register to 0, use 0 as the parameter value for the
command (for example, *SRE 0).
D-12
Series 2600 System SourceMeter® Instruments Reference Manual
Figure
D-7, the summary bits are logically ANDed (&) with the
"General bus commands"
Table
D-3. Note that the table includes both common commands and their
Return to
Section Topics
in
Section
11). Serial polling obtains the serial poll
and
"Reading registers"
2600S-901-01 Rev. C / January 2008
Table
D-3.
in this appendix.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

26362602261126122635

Table of Contents