Series 2600 System SourceMeter® Instruments Reference Manual
Model 2611/2612/2635/2636 safety shield
The maximum output voltage for a Model 2611/2612/2635/2636 channel is 200V, which is
considered hazardous and requires a safety shield
configuration in
to safety earth ground and chassis.
Figure 3-18
Safety shield for Models 2611/2612/2635/2636 hazardous voltage (200V maximum)
SMU 2611/
2612/2635/
2636
LO
Figure 3-19
Model 2601/2602-1 connections for test circuit shown in
2600S-901-01 Rev. C / January 2008
Figure 3-18
are shown in
Return to
Section Topics
(Figure
3-18). The connections for the test
Figure
3-20. Use #18AWG wire or larger for connections
Figure 3-18
Section 3: DUT Test Connections
3-17