Series 2600 System SourceMeter® Instruments Reference Manual
Figure 3-4
Model 2602/2612
Model 2636
WARNING
Sensing methods
Source-measure operations are performed using either 2-wire local sense connections or 4-wire
remote sense connections.
NOTE The default sense setting is 2-wire local. See
2600S-901-01 Rev. C / January 2008
Low-Noise Chassis Ground Banana Jack and Chassis Screw
Channel B LO
GND
Floating
1
FVR
Channel B LO
Chassis GND
When connecting to the model 2611, 2612, 2635 and 2636 SMU
outputs, with cables not rated for voltages above 42V, such as the
2600-ALG-2, you must disable the high voltage output by using the
INTERLOCK function as defined in section 10 of this manual.
Leaving the high voltage enabled while not properly insulating the
external connections to the unit poses a shock hazard which could
cause serious injury to the user. It is also recommended that the
LO connection terminal not be allowed to float by connecting it to
signal ground or another known signal reference.
later in this section to check and or change the sense mode.
Return to
Channel A LO
GND
Floating
Channel A LO
Section Topics
Section 3: DUT Test Connections
"Sense mode selection"
3-5