Keithley 2601 Reference Manual page 246

System sourcemeter 2600 series (smu)
Hide thumbs Also See for 2601:
Table of Contents

Advertisement

Section 10: Digital I/O and Triggering
Figure 10-16: SynchronousA output trigger
Output characteristics:
The trigger.assert command releases the line if the line is latched low
The physical line state does not change until all drives (internal and external) release the
line
Synchronous trigger mode
The Synchronous trigger mode is a combination of SynchronousA and SynchronousM trigger
modes. Use the Synchronous trigger mode for backwards firmware compatibility.
The SynchronousA and SynchronousM trigger modes provide additional flexibility. It is
recommended that you use SynchronousA and SynchronousM for firmware v1.4.0 or higher, and
use the Synchronous trigger mode for firmware prior to v1.4.0.
Figure 10-17: Synchronous input trigger
Input characteristics:
The falling edge generates an input event and latches the internal drive low
10-22