Keithley 2601 Reference Manual page 406

System sourcemeter 2600 series (smu)
Hide thumbs Also See for 2601:
Table of Contents

Advertisement

Section 13: Factory Scripts
Table 13-17
KIGeneral TSP test script: QueryPulseConfig
TSP project name:
TSP test script name:
Firmware version:
Function:
QueryPulseConfig
Usage:
tbl = QueryPulseConfig(tag)
Description
Parameters
Return
Values
Output
Data
13-28
KIFactoryPulse
KIPulse
1.2.0 and later
Once a pulse train has been configured and assigned to a tag, it is often desirable
to inspect the settings of this pre-configured pulse train. QueryPulseConfig()
can be used for this purpose. This function will return a table containing the
various settings associated with the tag input parameter.
tag: Numeric identifier of pulse train configuration being queried.
tostring():A function that returns most elements in a string convenient for
printing.
Identifying tag for this pulse train.
tag:
The smu configured for pulsing.
smu:
Pulse function: smuX.OUTPUT_DCAMPS or
func:
smuX.OUTPUT_DCVOLTS
Pulse bias level.
bias:
Pulse level for non sweeping pulses.
level:
Starting level for sweep pulses.
start:
Ending level for sweep pulses.
stop:
Limit value.
limit:
On time in seconds.
ton:
Off time in seconds.
toff:
The number of points in this pulse train.
points:
Reference to buffer containing measurement data
buf:
The sync_in digio line, if used.
sync_in:
The sync_out digio line, if used.
sync_out:
sourcevalues: A table containing the source value for each point in the pulse
train.
None
Return to
Section Topics
Series 2600 System SourceMeter® Instruments Reference Manual
2600S-901-01 Rev. C / January 2008

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

26362602261126122635

Table of Contents