Section 6: Range, Digits, Speed, Rel, and Filters

Overview

The documentation in this section provides detailed information on characteristics and script
programming for each of the following functions:

"Range"

"Digits"
"Speed"
"Rel"
"Filters"
Range
The selected measurement range affects the accuracy of the measurements as well as the
maximum signal that can be measured. Note that dashed lines are displayed (i.e., --.---- µA), to
indicate that the previous measurement is not recent. This usually happens when a change occurs
such as selecting a different range.

Available ranges

Table 6-1
2600 System SourceMeters
Table 6-1
Source and measurement ranges
Model 2601/2602
Voltage Ranges Current Ranges
100mV
1V
6V
40V
1. 10A range available only in pulse mode.
6-2
lists the available source and measurement ranges for the Keithley Instruments Series
®
.
Model 2611/2612
Voltage Ranges Current Ranges Voltage Ranges
100nA
200mV
1μA
2V
10μΑ
20V
100μΑ
200V
1mA
10mA
100mA
1A
3A
Return to
Series 2600 System SourceMeter® Instruments Reference Manual
100nA
1μA
10μΑ
100μΑ
1mA
10mA
100mA
1A
1.5A
1
10A
Section Topics
Model 2635/2636
Current Ranges
200mV
100pA
2V
1nA
20V
10nA
200V
100nA
1μA
10μΑ
100μΑ
1mA
10mA
100mA
1A
1.5A