Section 16: Calibration
v.
4.
Be sure to complete steps a through v for the 100nA to 1A ranges before continuing with 3A
and 10A range calibration (Model 2601/2602) or 1.5A and 10A range calibration (Model
2611/2612/2635/2636).
5.
Change connections as shown in
as shown).
6.
Select the DMM DC volts function.
7.
Repeat steps a through v for the 3A and 10A ranges (Model 2601/2602) or 1.5A and 10A
ranges (Model 2611/2612/2635/2636). Compute the current reading from the DMM voltage
reading and characterized 0.5Ω resistance value: I = V/R.
8.
Select auto polarity mode:
smua.cal.polarity = smua.CAL_AUTO
16-16
src_Z
DMM_Z_rdg
src_FS
DMM_FS_rdg
Typical values for the 1A range:
smua.source.calibratei(-1,-1e-10,-1e-5,-0.9,-0.892)
If this step is not on the CALA sense mode, send the measure calibration command
using the multimeter and Series 2600 readings, and range setting for the parameters:
smua.measure.calibratei(-range,Z_rdg,DMM_Z_rdg,
FS_rdg,DMM_FS_rdg)
Where: -range
Z_rdg
DMM_Z_rdg
FS_rdg
DMM_FS_rdg
Typical 1A range values:
smua.measure.calibratei(-1,-1e-4,-1e-5,-0.91,-0.892)
Return to
Section Topics
Series 2600 System SourceMeter® Instruments Reference Manual
-zero 2600 source output value
-zero DMM measurement
-FS 2600 source output value
-FS DMM measurement
negative of the present calibration range
-zero 2600 measurement
-zero DMM measurement
-FS 2600 measurement
-FS DMM measurement
Figure 16-3
(use 4-wire connections to the 0.5Ω resistor
2600S-901-01 Rev. C / January 2008