smuX.measure.delayfactor
Attribute
This attribute is a multiplier to smuX.DELAY_AUTO measurement delays.
Usage
delayfactor = smuX.measure.delayfactor
smuX.measure.delayfactor = delayfactor
Remarks
• The delay factor is only applied when
smuX.measure.delay = smuX.DELAY_AUTO.
• The default value is 1.0.
• This attribute can be set to a value less than 1 (e.g. 0.5) to decrease the automatic delay.
• This attribute can be set to a value greater than 1 (e.g. 1.5 or 2.0) to increase the
automatic delay.
• Setting this attribute to zero disables delays, even when
smuX.measure.delay = smuX.DELAY_AUTO.
Also see
smuX.measure.delay
Example
smuX.measure.delayfactor = 2.0
smuX.measure.filter.count
Attribute
Number of measured readings to yield one filtered measurement.
Usage
count = smuX.measure.filter.count
smuX.measure.filter.count = count
count
This attribute is the number of measurements that will be performed to yield one
Remarks
filtered measurement.
The
reset
Details
See "Filters" in
Also see
smuX.measure.filter.enable,
Sets filter count for SMU A:
Example
smua.measure.filter.count = 10
smuX.measure.filter.enable
Attribute
Enables/disables filtered measurements.
Usage
filter = smuX.measure.filter.enable
smuX.measure.filter.enable = filter
Set
filter
or
smuX.FILTER_OFF
0
or
smuX.FILTER_ON
1
This attribute enables or disables the filter.
Remarks
The
reset
• The median filter uses a first-in, first-out stack. When the stack (filter count) becomes full,
the "middle-most" reading is returned. For each subsequent conversion placed into the
stack, the oldest reading is discarded. The stack is then re-sorted, yielding a new
reading. If the filter count is an even number, the reading returned is the average of the
two middle readings.
Details
See "Filters" in
Also see
smuX.measure.filter.count,
• The median filter uses a first-in, first-out stack. When the stack (filter count) becomes full,
the "middle-most" reading is returned. For each subsequent conversion placed into the
stack, the oldest reading is discarded. The stack is then re-sorted, yielding a new
reading. If the filter count is an even number, the reading returned is the average of the
two middle readings.
2600S-901-01 Rev. C / January 2008
X= SMU channel (a or b)
(Models 2635 and 2636 only)
X = SMU channel (a or b)
function sets the filter count to 1.
Section
6.
smuX.measure.filter.type
X = SMU channel (a or b)
to one of the following values:
function disables the filter.
Section
6.
smuX.measure.filter.type
Return to
Section Topics
--Read the delay factor.
--Set the delay factor.
Increase the automatic delay by 2 times.
-- Reads filter count.
-- Writes filter count.
Set filter count from 1 to 100.
-- Reads on/off state of the filter.
-- Writes on/off state of the filter.
Disables the filter.
Enables the filter.