Section 13: Factory Scripts
Table 13-6 (continued)
KIGeneral TSP test script: SweepVLogMeasureI
TSP project name:
TSP test script name:
Firmware version:
Example
Table 13-7
KIGeneral TSP test script: SweepIListMeasureV
TSP project name:
TSP test script name:
Firmware version:
Function:
SweepIListMeasureV(smu, ilist, stime, points)
Description
Parameters
Data
Example
13-10
KIFactoryGeneral
KIGeneral
1.0.2 and later
SweepVLogMeasureI(smua, 1, 10, 0.001, 5)
This function performs a 5-point logarithmic voltage sweep starting at 1V and
stopping at 10V. Current is measured at every step (point) in the sweep. The
source will be allowed to settle on each step for 1ms before a measurement is
performed.
The following log values and corresponding source levels for the 5-point log
sweep are listed as follows:
MeasurePoint
LogStepSize
1
0.25
2
0.25
3
0.25
4
0.25
5
0.25
KIFactoryGeneral
KIGeneral
1.0.2 and later
Performs a current list sweep with voltage measured at every step (point):
• Sets the
to output the first
smu
settle for
seconds and then performs a voltage measurement.
stime
• Sets the
to output the next
smu
settle for
seconds and then performs a voltage measurement.
stime
• Repeats the above sequence until the voltage is measured for the last
amps value. The last point in the list to be measured is
smu, ilist, stime, points
SourceMeter Channel (A or B). Defaults to SMU A if all
smu:
parameters are omitted when function is called.
Arbitrary list of current source values:
ilist:
ilist = {value1, value2, ....valueN}
Settling time in seconds. Occurs after sourcing a value and before
stime:
performing a measurement.
Number of sweep measure points (must be
points:
Voltage measurements, current source values and timestamps are stored in
smuX.nvbuffer1.
myilist = {-100E-9, 100E-9, -1E-6, 1E-6, -1E-3, 1E-3}
SweepIListMeasureV(smua, myilist, 500E-6, 6)
This function performs a 6-point current list sweep starting at the first point in
myilist. Voltage is measured at every step (point) in the sweep. The source will
be allowed to settle on each value for 500µs before a measurement is performed.
Return to
Section Topics
Series 2600 System SourceMeter® Instruments Reference Manual
LogStep
SourceStepLevel
0.0
1.0000V
0.25
1.7783V
0.5
3.1623V
0.75
5.6234V
1.0
10.000V
amps value, allows the source to
ilist
amps value, allows the source to
ilist
≥ 2).
.
points
2600S-901-01 Rev. C / January 2008