Min And Max Chamber Temperature Limits; Cooler Temperature - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Min and Max Chamber
Temperature Limits

Cooler Temperature

Thermo Fisher Scientific
CHAMBER TEMPERATURE:
ACTUAL
>MIN
MAX
RANGE
AVG
The Minimum Chamber Temperature alarm limit screen is used to change
the minimum chamber temperature alarm limit. The minimum and
maximum chamber temperature screens function the same way.
In the Main Menu, choose Alarms > Chamber Temp > Min or Max.
CHAMBER TEMPERATURE:
ACTUAL MIN:
SET MIN TO:
RANGE
AVG
The Cooler Temperature screen displays the current cooler temperature
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from -40 to 10 °C. If the cooler temperature reading goes beyond
either the minimum or maximum limit, an alarm is activated. The word
"ALARM" appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Cooler Temp.
COOLER TEMPERATURE:
ACTUAL
>MIN
MAX
RANGE
AVG
o
49.0
C
o
47.0
C
o
51.0
C
DIAGS ALARM
o
47.0
C
o
48.0
C ?
INC/DEC
SAVE VALUE
DIAGS ALARM
o
-9.8
C
o
-20.0
C
o
-1.0
C
DIAGS ALARM
Model 42i Trace Level Instruction Manual
Operation
Alarms Menu
3-65