iCAP Customer Training Manual

Spray Chamber assembly

Start by connecting the drain tubing as figured below
Figure 11 – Spray chamber drain
Insert drain tube connector until it meets the main chamber.
Figure 12 – Nebuliser to Spray Chamber
The O-rings in the spray chamber should be inspected and replaced if any wear, or
damage, is visible. Push the Teflon insert with the attached sample tubing (supplied with
the nebuliser) into the rear of the nebuliser as far as possible without exerting undue
pressure.
Using a twisting motion insert the nebuliser into the spray chamber so that the collar is a
tight fit. The collar will set the insertion depth and aid reproducibility of results.