Range Mode Selection - Thermo Scientific 450iQ Instruction Manual

Hydrogen sulfide and sulfur dioxide analyzer
Table of Contents

Advertisement

Range Mode Selection

Thermo Scientific
The Range Mode Selection screen is used to switch between the various
range modes: Single, Dual, and Auto Range.
Home Screen>Settings>Measurement Settings>Range Mode Selection
The Range Mode Selection screen contains the following information:
Single: In single range mode, there is one range, one averaging time,
G
and one span coefficient.
Dual: In the dual range mode, there are two independent analog
G
outputs. These are labeled simply as the "High Range" and the "Low
Range". Each channel has its own analog output range setting,
averaging time, and span coefficient.
This enables the sample concentration reading to be sent to the analog
outputs at two different ranges. For example, the low SO
output can be set to output concentrations from 0 to 50 ppb and the
high SO
analog output set to output concentrations from 0 to 10000
2
ppb.
In addition, each SO
range can be calibrated separately. This is especially necessary if the two
ranges are not close to one another. For example, the low SO
set to 0–50 ppb and the high SO
Auto: The auto range mode switches the SO
G
outputs between high and low range settings, depending on the
concentration level. The high and low ranges are defined in the Range
Settings screen.
For example, suppose the low range is set to 50 ppb and the high range
is set to 100 ppb, as shown below. Sample concentrations below 50
ppb are output based on low range selection and sample concentrations
above 50 ppb are output based on high range selection. When the low
analog output has a span coefficient so that each
2
range is set to 0–10000 ppb.
2
Operation
Settings
analog
2
range is
2
, H
S, and CS analog
2
2
450iQ Instruction Manual
3-75

Advertisement

Table of Contents
loading

Table of Contents