Installation and Setup

Setup Procedure

Setup Procedure
2-6
42iQ Instruction Manual
Use the following procedure to setup the instrument:
1. Connect the sample line to the SAMPLE bulkhead on the rear panel
(Figure
2–5). Ensure that the sample line is not contaminated by dirty,
wet, or incompatible materials. All tubing should be constructed of
PTFE, 316 stainless steel, borosilicate glass, or similar tubing with an
OD of 1/4-inch and a minimum ID of 1/8-inch. The length of the
tubing should be less than 10 feet.
Note Gas must be delivered to the instrument free of particulates. It may
be necessary to use the PTFE particulate filter as described in
Particulate
Filter" on page 9-26.
Note Gas must be delivered to the instrument at atmospheric pressure. It
may be necessary to use an atmospheric bypass plumbing arrangement as
shown in
Figure 2–6
if gas pressure is greater than atmospheric pressure.
2. Connect the EXHAUST bulkhead to a suitable vent. The exhaust line
should be 1/4-inch OD with a minimum ID of 1/8-inch. The length of
the exhaust line should be less than 10 feet. Verify that there is no
restriction in this line.
3. If the optional zero/span solenoid valves are installed, connect a source
of NO
-free air to the ZERO IN bulkhead, and connect a source of
x
NO span gas to the SPAN bulkhead.
4. Connect a suitable recording device to the rear panel connector. For
detailed information about connecting to the instrument, refer to:
"Connecting External
Communications >
on page 3-62.
5. Plug the instrument into an outlet of the appropriate voltage and
frequency.
Note If instrument is equipped with an internal permeation oven, refer to
Chapter 9,
"Optional
Equipment" for setup instructions.
Devices" on page 9-1
"Analog
I/O" on page
"PTFE
3-60
and
"Digital
I/O"